+8618117273997weixin
English
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語

Damaged area and ignition times 45°test machine

Product No: LS-GBT14645

Leave a Message

=
  • Description
  • The LS-GBT14645 Damaged area and ignition times 45°test machine is professionally designed and manufactured by LISUN in accordance with the requirements of GB/T 14645. It is dedicated to the 45° test for determining the damaged area and ignition times of textiles, providing accurate and reliable test data for the flame retardant performance evaluation of textile products, and helping related products meet the relevant national and international standard requirements.

    Standards:

    Standards No. Standards Name
    GB/T14645-2014 纺织织物-燃烧性能-45°方向损毀面积和接焰次数测定
    ASTM D1230 Standard Test Method for Flammability of Apparel Textiles

    Specifications:

    Index Parameter
    Dimension 1200(L) × 500(W) × 700(H)mm
    Power Supply AC 220V, 50/60Hz, 5A
    Applied Gas ≥95% Propane (to be prepared by the customer)
    Weight About 50kg

    Main Features
    • Equipped with heat resistant glass door for convenient observation of the internal test state;
    • The internal tooling is made of stainless steel, ensuring good corrosion resistance and service life;
    • The top and side of the cabinet are designed with 6 ventilation holes to ensure the air circulation in the test chamber;
    • The sample holder needs to be replaced for different types of experiments to meet diverse test requirements;
    • Adopt imported flow control valve to precisely control the flame height;
    • Equipped with imported pressure gauge and pressure regulating valve to clearly display the gas pressure in real time.

    Damaged Area Test Function
    • The flame height can be adjusted by the needle valve, with the standard flame height of 45±2mm;
    • Imported pressure gauge and pressure regulating valve can display gas pressure to ensure stable gas supply;
    • The planimeter for measuring the damaged area is optional, facilitating the quantitative analysis of test results.

    Ignition Times Test Function
    • The sample holder is placed at a 45° angle, in line with the standard test posture requirements;
    • The timer has an accuracy of 0.1s and supports automatic timing, ensuring the accuracy of time-related test data;
    • The sample holder is made of 0.5mm hard stainless steel wire wound into a spiral ring with an inner diameter of 10mm, with stable structure and high matching degree with the test standard;
    • The burner can use butane or propane gas, with a gas outlet diameter of 0.2mm and a nozzle length of 6.4mm;
    • The gas flow range is 0-1000ml/min, which can be adjusted according to test needs;
    • Support electronic ignition, and the test time is displayed digitally for intuitive reading.