Product No: LS-B48
The LS-B48 Oxygen Bomb Air Aging Test Chamber is designed and manufactured in full compliance with IEC 60811-1-2 Clause 8.2 and 8.3 (equivalent to GB/T 2951.12 Clauses 8.2 and 8.3), IEC 60335-1 Clause 22.32 (equivalent to GB/T 4706.1 Clause 22.32), as well as GB/T 5013.1 and IEC 60245-1. It is suitable for aging tests of medical device materials, wires and cables, rubber (plastic, artificial leather) insulation layers and sheaths, and other polymer materials in oxygen and compressed air media under specified pressure and temperature conditions.
Specification:
Air Bomb Test | The medium is air, the temperature is 127℃, and the working pressure is 0.55±0.02 MPa (in accordance with standards such as IEC 60811-1-2 Clause 8.2). |
Oxygen Bomb Test | The medium is oxygen with a purity of not less than 97%, the temperature is 70℃, and the working pressure is 2.1±0.07 MPa (in accordance with standards such as IEC 60811-1-2 Clause 8.3). |
Working Pressure | ≤2.5MPa |
Pressure Display Accuracy | 0.02MPa |
Temperature Automatic Control Range | Room temperature~150℃ |
Temperature Deviation | ±0.5℃ |
Cabinet Size | 840*420*880mm |
Tank Volume | ≥4000cm³ |
Pressure Reduction Program | After the test, the test chamber automatically relieves pressure gradually, with the pressure relief time not less than 5 minutes until the pressure drops to atmospheric pressure. |
Built-in Timer | 0.01s~999.9s/m/h Arbitrary Setting |
Power Supply Voltage | AC220~240V 50/60Hz |
Air/Oxygen Source | Customer Prepared: Compressed air and oxygen with a purity of not less than 97% |
IEC 60811-1-2 Clause 8.2 and 8.3