ISO 7176-21:2025 Wheelchairs — Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers

ISO 7176-21:2025 specifies electromagnetic compatibility (EMC) requirements and test methods for electrically powered wheelchairs, scooters and their associated battery chargers. It defines a set of immunity tests against common electromagnetic disturbances, including electrostatic discharge (ESD), radiated radio-frequency fields, conducted radio-frequency disturbances, voltage dips and short interruptions. Specialised test provisions are tailored for wheelchair products, such as the charged-frame ESD test mode and table-top configuration for radiated RF immunity testing, with strict safety-oriented performance criteria to prevent unintended movement of drive wheels, accidental release of automatic brakes and unwanted activation of safety-related auxiliary electric components during interference.

This standard also mandates an additional proximity field immunity test for RF wireless communication equipment following IEC 60601-1-2:2020 Table 9. Test setups, frequency steps, dwell time, sample orientations and recovery rules for battery chargers are clearly specified. All evaluation aims to ensure that powered mobility devices maintain safe functions in real electromagnetic environments and avoid hazards for occupants when exposed to electromagnetic interference.

LISUN supplies the fully test solutions for the following ISO 7176-21:2025, please see the below excel table:

ISO 7176-21:2025 Clause and Subject LISUN MODEL
5.2.2 Electrostatic discharge(ESD) immunity ESD61000-2
5.3.5/10.1 Electrostatic discharge immunity
5.2.3/5.3.6 Radiated radio frequency field immunity RFRI61000-3
5.2.4/10.6 Power frequency magnetic field immunity PFM61000-8A
5.3.1 Mains terminal disturbances EMI-9KB
5.3.2 Radiated emissions
5.3.3 Harmonic current emissions LSHF61000-3-2/3
5.3.4 Voltage fluctuations and flicker
5.3.6/10.2 Radiated radio frequency field immunity AC-3M
5.3.7/10.3 Fast transient/burst immunity EFT61000-4
5.3.8/10.4 Surge immunity SG61000-5
5.3.9/10.5 Conducted disturbance immunity RFCI61000-6
5.3.10/10.7 Voltage dips and short interruptions immunity CSS61000-11
10.2.1.3 GTEM cell test method GTEM-2