Product No: RFCI61000-6

  • Specifications
  • Videos
  • Downloads
  • Applications
  • Standards
  • Installations
  • Tech Articles
  • The RFCI61000-6 series RF Conducted Immunity Test System is used to simulate conducted interference signals induced by radio frequency (RF) fields. These signals are injected into the power and signal lines of the equipment under test (EUT) using standardized injection methods (CDN, electromagnetic clamp, and current clamp). This method evaluates the EUT’s operational stability in the presence of such conducted RF interference.

    The RFCI61000-6 series Conducted RF Immunity Testing System innovatively integrates three core modules: a signal source, power amplifier and power meter. It is also compatible with LISUN’s full range of coupling-decoupling networks (CDN), electromagnetic injection clamps and current injection clamps, providing flexible adaptation to the testing needs of different EUT types. It equipped with a large Android touch screen display with English and Chinese support, the system is widely used for EMC compliance testing in a variety of fields, including LED lighting, medical, power, industrial control, new energy charging stations and communications.

    Standards:

    Standards No. Standards Name
    IEC 61000-4-6:2023 Electromagnetic compatibility (EMC) – Part 4-6: Testing and measurement techniques – Immunity to conducted disturbances, induced by radio-frequency fields
    EN 61000-4-6:2023 Electromagnetic compatibility (EMC) – Part 4-6: Testing and measurement techniques – Immunity to conducted disturbances, induced by radio-frequency fields
    GB/T 17626.6-2017 电磁兼容 试验和测量技术 射频场感应的传导骚扰抗扰度

    Specifications:

    Parameters LISUN Model RFCI61000-6-35W RFCI61000-6-85W
    1. Output performance Output frequency range 150kHz~230MHz 150kHz~230MHz
    Output power 35W 85W
    Output voltage
    (CDN Method)
    1~17Vrms 1~30Vrms
    Output voltage
    (Electromagnetic Clamp Method)
    1~11Vrms 1~18Vrms
    Output voltage
    (Current Clamp Method)
    N/A 1~10Vrms
    Output impedance 50Ω(±5%)
    Voltage Standing Wave Ratio
    (VSWR)
    ≤1.2
    2. Built-in signal source Harmonic suppression <15dBc
    Frequency range 9kHz~1GHz
    Output Level -60~10dBm
    Signal Type Continuous Wave (CW)
    Amplitude Modulation Modulation frequency 1Hz~10kHz, modulation depth 1%~99%
    3. Built-in power amplifier Pulse modulation Modulation frequency 1Hz~1kHz, duty cycle 1%~100%
    Maximum output power +44dBm(35W) +49dBm(85W)
    4. Built-in 3 sets power meter (Monitoring, Forward and Reverse Power Meter) LISUN Model LS-PM9K400M-MN、LS-PM9K400M-FW、LS-PM9K400M-RV
    Frequency range 9kHz~400MHz
    Dynamic Range -60dBm~20dBm

    Instructions:
    Please refer to Figure 20 of Section 7.4.2 of IEC 61000-4-6:2023 or GB/T 17626.6-2017 to select the following injection method: CDN method, electromagnetic clamp method, or current clamp method.

    Figure 20 – Rules for selecting the injection method

    Recommendation 1:
    The RFCI61000-6 series can be used with the following LISUN CDN coupling and decoupling networks to perform CDN injection testing on the EUT (Please refer to IEC 61000-4-6:2023 Clause 6.2.2 and Table 4 to select the right CDN to test):

    CDN for RF Conducted Immunity Test

    CDN Types LISUN Model Test Ports Power
    Power Supply Cables CDN-M1-16 Single -phase single-wire power cable (L) 220V/16A
    CDN-M1-32 Single -phase single-wire power cable (L) 220V/32A
    CDN-M2-16 Single -phase two-wire power cable (L/N) 220V/16A
    CDN-M2-32 Single -phase two-wire power cable (L/N) 220V/32A
    CDN-M2-100 Single -phase two-wire power cable (L/N) 220V/100A
    CDN-M3-16 Single -phase three-wire power cable (L/N/PE) 220V/16A
    CDN-M3-32 Single -phase three-wire power cable (L/N/PE) 220V/32A
    CDN-M3-100 Single -phase three-wire power cable (L/N/PE) 220V/100A
    CDN-M3-32-3 Three -phase and three-wire power cable (3L) 380V/32A
    CDN-M4-32 Three -phase and four-wire power cable (3L/1N) 380V/32A
    CDN-M4-100 Three -phase and four-wire power cable (3L/1N) 380V/100A
    CDN-M4-32-PE Three -phase and four-wire power cable (3L/1PE) 380V/32A
    CDN-M5-32 Three -phase and five-wire power cable (3L/1N/1PE) 380V/32A
    CDN-M5-100 Three -phase and five-wire power cable (3L/1N/1PE) 380V/100A
    Unscreened Balanced Lines CDN-T2 2 line 100V/5A
    CDN-T4 4 line 100V/5A
    CDN-T8 8 line with RJ45 port 100V/5A
    Unscreened Unbalanced Lines CDN-AF2 2 line 100V/5A
    CDN-AF4 4 line 100V/5A
    CDN-AF8 8 line 100V/5A
    Screened Cables CDN-S1 single line with BNC port 100V/5A
    CDN-S4 4 line with USB ports 100V/5A
    CDN-S8 8 line with RJ45 port 100V/5A
    CDN calibrating jigs CDN-CS100 Including 150Ω-50Ω adapter resistors. Note: If one or more CDNs are purchased, one set of CDN-CS100 must be selected.

    Recommendation 2:
    The RFCI61000-6 series can be used with the LISUN AB-EM electromagnetic injection clamp to perform electromagnetic injection testing on the EUT, which is fully according to IEC 61000-4-6:2023 Clause 6.2.3.3.

    AB-EM Electromagnetic Clamp for RF Conducted Immunity Test

    AB-EM Electromagnetic Clamp for RF Conducted Immunity Test

    • Injection power: Maximum 100W (5 minutes)
    • Injection frequency: 150kHz~400MHz
    • Coupling/decoupling directivity: Greater than 10dB (>10MHz)

    Recommendation 3:
    The RFCI61000-6 series can be used with the LISUN AB-100W current injection clamp to test signal lines of any specifications, which is fully according to IEC 61000-4-6:2023 Clause 6.2.3.2.

    AB-100W Current Injection Clamp for RF Conducted Immunity Test

    AB-100W Current Injection Clamp for RF Conducted Immunity Test

    • Supports testing of any signal cable specification: unshielded balanced cables (T2, T4, T8), unshielded unbalanced cables (AF2, AF4, AF8), communication cables (RJ11, RJ11S, RJ45, RJ45S, USB-C, USB-P), and any other cable specification
    • Maximum supported cable diameter: 40mm
    • Maximum output power: 100W
    • Frequency range: 100kHz to 300MHz
    • Insertion loss: <7dB
    • Includes standard current clamp calibration kits

    Tags:
  • EN61000-4-6 2014 Standard Free Download
  • Motor-Operated Tool Test Solutions

    LISUN's Motor-Operated Tool | Power Tool Testing solutions strictly comply with a range of core...

    Electric Toys Test Solutions

    LISUN's electric toys test solutions are engineered to meet the stringent requirements of key...

    Transformer Test Solutions

    LISUN's transformer test solutions are designed to meet the comprehensive requirements of...

    Energy Meter Test Solutions

    LISUN's energy meter test solutions are designed to ensure the accuracy and reliability of...

  • IEC 60255-1:2022 establishes the general technical requirements for measuring relays and protection equipment used in power systems. It defines basic terms, rated values, construction, safety,...
    IS 16102 (Part 2): 2026 specifies the performance requirements for self‑ballasted LED lamps used for general lighting services in India, complementing the safety provisions defined in Part 1. This...
    CISPR 14-2:2020 is a product family standard under the International Special Committee on Radio Interference (CISPR), dedicated to the electromagnetic immunity requirements of household appliances,...
    As a revised edition released in 2023, IEC 61869-1:2023 supersedes the 2007 version of the same standard as well as IEC 61869-6:2016, updating technical content to meet the latest industry needs of...
    IEC 60730-1:2010 defines general safety and performance requirements for automatic electrical controls used in or with household and similar equipment, encompassing energy sources like electricity,...
    UL 1447:2020 is a safety standard developed by Underwriters Laboratories (UL) for electric lawn mowers, covering both corded (rated voltage ≤250V) and battery-powered models. It specifies...
    IEC 62841-1:2014 "Electric Motor-Operated Hand-Held, Transportable Tools and Lawn and Garden Machinery - Safety - Part 1: General Requirements" establishes foundational safety guidelines for handheld...
    IEC 62115:2017 Electric toys—Safety is an international standard established by the International Electrotechnical Commission (IEC), specifying safety requirements for electric toys designed or...
    IEC 62041:2010 sets forth the electromagnetic compatibility (EMC) requirements for transformers, power supplies, reactors, and similar products. This standard covers a wide range of equipment,...
    The IEC 61547:2020 standard specifies the electromagnetic compatibility (EMC) immunity requirements for equipment intended for general lighting purposes. It applies to lighting devices such as LED...
  • Indonesia – LISUN support customer to set up EMI/EMC test lab

    21 Oct, 2024

    This Indonesia-based product certification body and testing laboratory is one of the leading quality testing and...

    Indonesia – LISUN engineer provide installation and training for EMI/EMC test equipment

    10 Mar, 2025

    In the field of testing, certification, and inspection, this testing laboratory has consistently proven to be a...

    South Korea – Customer Successfully Installed and Used LISUN EMC Electromagnetic Compatibility Testing Equipment

    20 Jul, 2025

    The customer in this case is a leading high-quality sensor manufacturer in South Korea, specializing in the field of...

    Russia – LISUN Engineer Provide Installation and Training for EMI and EMC Testing Equipments

    10 Aug, 2025

    As a leading benchmark in Russia’s lighting industry, this enterprise is positioned around providing...

    The In-depth Cooperation between Russia’s Leading Lighting Enterprise and Professional Testing Equipment Supplier

    10 Apr, 2026

    I. Company Overview: A Benchmark in Full-Scene Lighting Solutions As a leading lighting enterprise and prominent LED...

  • Explain RF conducted immunity test system

    26 Dec, 2022

    One of the electromagnetic compatibility tests is a continuously performed radio­frequency (RF)...

    Exploring What is RF Immunity Test: The Significance of RF Conducted Immunity Testing System

    16 Oct, 2023

    What is RF Immunity Test: RF Conducted Immunity Testing is a common electromagnetic compatibility...

    What is RF Immunity Test? An In-depth Analysis of the RF Conducted Immunity Test System Based on the LISUN RFCI61000-6

    22 May, 2025

    Abstract As the complexity of electronic devices continues to increase, ensuring electromagnetic...

    RF Conducted Immunity Test System: A Detailed Overview of LISUN RFCI61000-6

    15 Apr, 2025

    Introduction The RF Conducted Immunity Test System is an essential component in the field of...

    Immunity to Conducted Disturbances: LISUN RFCI61000-6 RF Conducted Immunity Test System

    02 Jun, 2025

    Abstract The RF conducted immunity test system is a crucial tool for assessing the immunity to...