Product No: EMI-9KA
R&S®ESR is an EMI test receiver compliant with the CISPR 16-1-1 standard. It supports both traditional stepped-frequency scans and optional FFT-based time-domain scans. Equipped with built-in preselection filters, a 20 dB preamplifier, and a highly linear front end, it offers an excellent balance of high dynamic range and high sensitivity. An optional real-time spectrum analysis feature—with a bandwidth of up to 40 MHz—includes spectrograms, persistence displays, and frequency-mask triggering, enabling the capture of elusive or sporadic emissions. The R&S®ESR is widely used for CE (conducted emission) and RE (radiated emission) certification testing across various sectors, including third-party certification laboratories, automotive electronics, new energy, consumer electronics and home appliances, industrial automation, rail transport, medical devices, IoT communications, lighting, and the defense and aerospace industries. For more product information, please click: R&S®ESR product brochure
Specifications:
| LISUN Model | ESR3 | ESR7 | ESR26 |
| Standard Frequency Range | 9 kHz ~ 3.6 GHz | 9 kHz ~ 7 GHz | 9 kHz ~ 26.5 GHz |
| Test Noise (preamplifier off) | 30 MHz ~ 1 GHz:< 8 dBµV; 1 GHz ~ 3.6 GHz:< 20 dBµV |
30 MHz ~ 1 GHz:< 8 dBµV; 1 GHz ~ 3.6 GHz:< 20 dBµV; 3.6 ~ 7 GHz:< 26 dBµV |
30 MHz ~ 1 GHz:< 8 dBµV; 1 GHz ~ 3.6 GHz:< 20 dBµV; 3.6 ~ 26.5 GHz:< 26 dBµV |
| Test Noise (preamplifier enabled) | 30 MHz ~ 1 GHz:< -3 dBµV; 1 GHz ~ 3.6 GHz:< 9 dBµV |
30 MHz ~ 1 GHz:< -3 dBµV; 1 GHz ~ 3.6 GHz:< 9 dBµV; 3.6 ~ 7 GHz:< 13 dBµV |
30 MHz ~ 1 GHz:< -3 dBµV; 1 GHz ~ 3.6 GHz:< 9 dBµV; 3.6 ~ 26.5 GHz:< 13 dBµV |
| Measurement Uncertainty | 9 kHz ~ 3.6 GHz:0.47 dB | 9 kHz ~ 3.6 GHz:0.47 dB; 3.6 GHz ~ 7 GHz:0.59 dB |
9 kHz ~ 3.6 GHz:0.47 dB; 3.6 ~ 7 GHz:0.59 dB; 7 ~ 13.6 GHz:1.01 dB; 13.6 ~ 26.5 GHz:1.34 dB |
| Preselector | 16 fixed filters; analyzer mode can be disabled | ||
| Preamplifier | 1 kHz – 7 GHz, 20 dB gain, switchable | ||
| Resolution Bandwidth | 10 Hz – 10 MHz (–3 dB), in 1/2/3/5/10 steps; CISPR: 200 Hz, 9 kHz, 120 kHz (–6 dB), 1 MHz (impulse bandwidth) | ||
| Detector | Maximum peak, minimum peak, quasi-peak, RMS, average, average with meter time constant (CISPR-average), and RMS average (CISPR-RMS) | ||
| 1 dB Compression Point | +5 dBm (nominal, 0 dB attenuation, preamplifier and preselector off) | ||
| Maximum Continuous-wave Level | 30 dBm (1 W) (attenuation ≥10 dB, preamplifier off) | ||
| Maximum continuous-wave level | 23 dBm (0.2 W) (attenuation ≥10 dB, preamplifier on) | ||
| Maximum Pulse Voltage | 150 V / 450 V (Input 1 / Input 2, attenuation ≥10 dB) | ||
| Maximum Pulse Energy | 1 mWs / 20 mWs (Input 1 / Input 2, attenuation ≥10 dB, 10 µs) | ||
| Display Screen | 21 cm (8.4″) touchscreen, 800 × 600 pixels, displays up to 6 traces on a single screen | ||
| Maximum Number of Scan Points | 4,000,000 | ||
| Remote Control Interface | GPIB(IEC 625-2 / IEEE 488.2); LAN(10/100/1000BASE-T); Provides LabVIEW, LabWindows/CVI, and VXIplug&play drivers | ||
| Compatible Standards | CISPR 16-1-1; other standards: CISPR, EN, ETS, ANSI, FCC, VCCI, MIL-STD-461 | ||
| More Detailed Specifications | Please click here to view: R&S®ESR product specifications | ||