3 Phases Digital Power Meter
User's Manual
Applicable LISUN models: LS8930
1. Basic Understanding of Equipment
1.1 Key Technical Specifications
| Parameters | Note |
| Voltage Range | 1000 V (Settings: 1000 V, 600 V, 300 V, 150 V, 60 V, 30 V) |
| Current Range | 50A (Settings: 50A, 20A, 10A, 5A, 2A, 1A) |
| Input Impedance | Voltage input impedance is approximately 2 MΩ; low-range current input impedance is approximately 4 mΩ. |
| Measurable Voltage Parameters | Urms: True RMS; Umn: Rectified average calibrated to RMS; Udc: Simple average; Urmn: Rectified average; Uac: AC component |
| Measurable Current Parameters | Irms: True RMS; Imn: Rectified average value calibrated to RMS; Idc: Simple average; Irmn: Rectified average; Iac: AC component |
| Other parameters that can be measured | Cumulative energy, cumulative ampere-hours, energy time, harmonics, active power, apparent power, reactive power, power factor, phase difference, frequency, U+pk, U-pk, I+pk, I-pk, P+pk, P-pk, peak factor, WP, WP+, WP-, q, q+, q- … |
1.2 Equipment Configuration Verification
1.2.1 Unpacking
When unpacking the device, handle it gently to avoid scratching the casing with sharp tools. After unpacking, first inspect the exterior of the device to ensure there is no visible deformation or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.
1.2.2 Main Unit and Accessories
| Number | Note |
| ① | LS8930 Host |
| ② | Power Cord |
| ③ | USB Cable |
| ④ | RS-485 communication cable, spare |
| ⑤ | RS-232 communication cable, spare |

1.2.3 Downloading Electronic Documents
We have already sent download links for the electronic user manual, software, warranty card, and other materials via email at the time of shipment. Please download them as soon as possible. If you missed the email, please feel free to contact us to request the download links again.
2. Safety Precautions
2.1 Electrical Safety
Before connecting the power cord, verify that the supply voltage is between AC 100 V and 240 V, with a frequency of 50 Hz or 60 Hz. You must use a three-prong outlet with a protective ground, and the protective ground resistance must be ≤4 Ω. Never cut or disconnect the protective ground wire, as doing so poses a risk of electric shock.
2.2 Environmental Safety
Do not use in environments with flammable or explosive substances, strong electromagnetic interference, high temperatures and humidity, or excessive dust levels. Operating environment: Temperature 5°C to 35°C, relative humidity ≤ 65% RH.
2.3 Operational Safety
Before wiring, the power to both the meter and the device under test must be turned off; wiring while the circuit is energized is strictly prohibited. Connect voltage wires in parallel and current wires in series; under no circumstances should a current circuit be connected to a voltage terminal or a voltage circuit to a current terminal. After wiring, tighten the terminals and ensure that no bare wires are exposed to prevent short circuits or electric shock.
Do not apply voltages or currents that exceed the rated range.
3. Equipment Overview and Usage
3.1 Introduction to the Back Panel and Wiring
3.1.1 Introduction to the Back Panel
| Number | Note |
| ① | Power Cord Jack |
| ② | Input Unit 1 Current Input Terminal |
| ③ | Input Unit 1 Voltage Input Terminal |
| ④ | Input Unit 2 Current Input Terminal |
| ⑤ | Voltage Input Terminal of Input Unit 2 |
| ⑥ | Input Unit 3 Current Input Terminal |
| ⑦ | Voltage Input Terminal of Input Unit 3 |
| ⑧ | RS232/RS485 interface, spare |
| ⑨ | USB port |
| ⑩ | Ethernet port, spare |

3.1.2 Input Module Wiring Diagram
Single-phase, two-wire (1P2W), connected to input unit 1:

Three-phase, four-wire (3P4W), connected to input terminals 1, 2, and 3:

Three Voltages, Three Currents (3V3A):

Three-phase, three-wire (3P3W), connected to input terminals 1 and 3:

Single-phase, three-wire (1P3W), connected to input terminals 1 and 3:

3.2 Introduction to the Front Panel
| Number | Note |
| ① | The wiring configuration display shows the default three-phase, four-wire system. This must match the actual wiring configuration on the back panel. Press the ② wiring button, then use ③ to select the correct wiring configuration, and finally press the ④ confirm button to confirm. |
| ⑤ | The three windows on the left display three-phase voltage by default; you can switch to display other parameters using the “Unit” and “Function” buttons at position ⑥. |
| ⑦ | By default, the four windows on the right display the three-phase current and total power. You can switch to display other parameters using the unit and function keys located to the right of each window, as shown at ⑧. |

4. Using the Software
4.1 Preparations
Connect the device to the computer using a USB cable. Open Device Manager on your computer; if you see the following ports listed, communication is working properly.

4.2 Software Overview
4.2.1 Software Startup Screen
| Number | Note |
| ① | Select the correct port number based on the ports displayed in Device Manager. |
| ② | Tap to start recording data, then tap again to stop recording. |
| ③ | You can export all data from the current record to an Excel spreadsheet. |
| ④ | Harmonic Features |
| ⑤ | Data Calculation Features |
| ⑥ | Exit the Software |

4.2.2 Harmonic Function Interface
| Number | Note |
| ① | Click to Read Harmonic Data |
| ② | Progress Bar |
| ③ | Message: Read successful |
| ④ | You can export the current harmonic data to an Excel spreadsheet. |
| ⑤ | Return to the Home Screen |

4.2.3 Data Calculation Interface
| Number | Note |
| ① | Select “Clear,” press the Settings button, and past data will be automatically cleared. |
| ② | Settings button |
| ③ | After clearing is successful, the mode automatically switches to Start; tap the Settings button again to automatically switch to Stop. |
| ④ | The device begins recording data |
| ⑤ | To stop, tap “Settings” again. The device will record and calculate the relevant data for this period. |

| Number | Note |
| ① | Click to retrieve the relevant data into the software |
| ② | Retrieval successful |
| ③ | You can export the current data to an Excel spreadsheet. |
| ④ | Return to the Home Screen |

5. Daily Maintenance and Care of Equipment
5.1 Routine Maintenance
5.1.1 Daily Cleaning
Be sure to turn off the power and unplug the power cord before cleaning. We recommend wiping the unit with a soft, antistatic cotton cloth and a mild detergent; do not spray liquids directly onto the unit. Do not use chemical solvents such as benzene, toluene, or acetone.
5.1.2 Routine Inspections
Before each use, check that the power cord and terminal connections are secure and free of burn marks; ensure that the ventilation openings are not blocked; and after turning on the unit, verify that the indicator lights and display are functioning properly.
5.2 Storage Specifications
For long-term storage, keep the product in a dry, dust-free, well-ventilated, and non-corrosive environment at a temperature of 5°C to 35°C and a relative humidity of ≤65% RH.
Turn off the power and unplug all cables; clean the equipment and wrap it in dust-proof film; turn on the power for 10 minutes each month to prevent moisture damage to the components.
6. Troubleshooting and Resolving Common Problems
| Fault Symptoms | Possible Causes | Procedure |
| Large Deviation in Measurement Values | Wiring error; the actual wiring configuration does not match the wiring configuration selected on the device’s front panel | Be sure to refer to Section 3.1.2 for the wiring scheme; be sure to refer to Section 3.2 when selecting the front panel wiring scheme. |
| Button not working | Keyboard is locked | Press SHIFT+HARMO (i.e., KEY PROT) to unlock the keyboard |
7. Packing List
| No. | Name | Quantity |
| ① | LS8930 Host | 1 unit |
| ② | Power Cord | 1 piece |
| ③ | USB Cable | 1 piece |
| ④ | RS-485 Communication Cable | 1 piece |
| ⑤ | RS-232 Communication Cable | 1 piece |
Appendix A: Indicator Lights and Label Descriptions
Unit Indicator Light Descriptions
The m (mill), k (kilo), and M (giga) indicators are order-of-magnitude indicators: 1M = 1000k, 1k = 1000, 1m = 0.001
| Indicator light | Parameters | Unit |
| V | Voltage | volt |
| A | Current | Ampere |
| W | Active Power | Tile |
| VA | Apparent Power | volt-ampere |
| Var | Reactive power | Tired |
| Hz | Frequency | Hertz |
| PF | Power Factor | |
| TIME | Time | Hour: Minute: Second |
| pk | Can be combined to form a peak | Determined by the group |
| h/h+ | Can be combined to form kilowatt-hours | Time |
| % | Relative Harmonic Content | |
| o | Phase difference | Degree |
| 1 | Unit 1 | |
| 2 | Unit 2 | |
| 3 | Unit 3 | |
| Σ | Integrated Unit |
Status Indicator Light
| Indicator light | Note | Notes |
| UPDATE | Operating Status Indicator Light | A steady light indicates normal operation; a flashing light indicates that data is being updated. |
| STORE | Storage Operation Indicator Light | A steady light indicates that the device is saving data; a flashing light indicates that data has been saved once. |
| SCALING | Magnification On Indicator Light | When the light is on, it indicates that the magnification is enabled; use the magnification data for calculations. |
| RMS | Mode Indicator Light | A lit indicator indicates that the calculated data is the effective value of the signal. |
| AC | Mode Indicator Light | When the light is on, it indicates that the calculated data represents the AC component of the signal. |
| DC | Mode Indicator Light | A lit indicator shows that the calculated data is the simple average of the signals. |
| V-MEAN | Mode Indicator Light | The voltage calculation data indicated by the lit light represents the rectified average value calibrated to the effective value. |
| KEYPROT | Button Protection | A lit indicator means the button is locked and will not respond. |
| AVG | Average Data Indicator | A lit indicator means that the calculated value is the average of multiple data points. |
| VOLTAGE | Voltage Range Indicator Light | |
| CURRENT | Current Range Indicator Light | |
| AUTO | Voltage or Current Auto-Range Indicator Light | When the light is on, it indicates that the voltage or current is within the range and is being automatically detected. |
| 3P4W | Wire-type indicator light | The indicator light shows that the meter is a three-phase, four-wire system. |
| 3V3A | Wire-type indicator light | The “Light On” Indicator Instrument Uses the 3-Voltage, 3-Current Meter Method |
| 3P3W | Wire-type indicator light | The indicator light shows that the meter is a three-phase, three-wire system. |
| 1P3W | Wire-type indicator light | The indicator light shows that the meter is single-phase, 3-wire. |
Button Indicator Light
| Indicator light | Note | Notes |
| MAXHOLD | Maximum Value Indicator | When the light is on, it indicates that some of the calculated data values are at their maximum. |
| HOLD | Data Retention | A lit light indicates that the calculated data is held, meaning the data is no longer being updated. |
| START | Points Accumulation Start Indicator Light | When the light comes on, it indicates that the accumulation of points has begun. |
| STOP | Point Accumulation Stop Indicator Light | When the light comes on, it indicates that integration has stopped. |
| L-FILTER | Line Filter Indicator Light | A lit light indicates that the filter is on. |
| F-FILTER | Frequency Filter Indicator Light | A lit light indicates that the filter is on. |
| HARMO | Harmonic Display On Indicator Light | The light is on, indicating that the harmonic display is on. |
| SHIFT | Second Function Indicator Light | The light is on, indicating that the second function is active. |