Pendulum Impact Test Device

User's Manual

Applicable LISUN models: IK01-06P

1. Basic Understanding of Equipment

1.1 Key Parameters

Parameters Note
Combined Mass of Impact Components 150±1 g and 200±1 g
Quality of Mounting Brackets 10 ± 1 kg
Bracket Rotation Angle ±60°
Swing Arm Length 1000 ± 1 mm
Swing Arm Quality 100 g ± 1 g
Material for the swing arm Steel pipe with an outer diameter of 9 mm and a wall thickness of 0.5 mm
Sensor Resolution Angle: 0.36 degrees; Height: 0.1 mm
Impact Energy 0.14 J, 0.2 J, 0.35 J, 0.5 J, 0.7 J, 1 J

Note: According to the requirements of the IEC 60068-2-75 (GB/T 2423.55) standard and the method for calculating equivalent mass, the previous version of the standard used a 150-g impact element, whose equivalent mass was 150 g + (100 g/2) = 200 g; The new standard uses a 200 g impact element, with an equivalent mass of 200 g + (100 g/2) = 250 g. The IK01-06P includes impact elements of both masses.

See Figure 1-1 for the external dimensions.

Pendulum Impact Test Device-Figure1

Figure 1-1

1.2 Equipment Configuration Verification

1.2.1 Unpacking

When unpacking the device, handle it gently to avoid scratching the casing with sharp tools. After unpacking, first inspect the exterior of the device to ensure there is no visible deformation or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, please contact us immediately.

1.2.2 Main Unit and Accessories

Main Unit: 1 IK01-06 Impact Element Impact Test System main unit.

Main Components (Figs. 1–2):

Number Note
Expansion anchors, used to mount the main unit to a vertical wall
USB charging cable, used to charge the display
Spare 150g impact element; the 200g impact element is already installed on the swing arm at the factory.

Pendulum Impact Test Device-Figure2

Figures 1–2

1.2.3 Downloading Electronic Documents

We have already sent the download links for the digital user manual, warranty card, and other documents via email at the time of shipment. Please download them as soon as possible. If you missed the email, you can contact us to request the download links again.

2. Safety Precautions

2.1 Operational Safety

You must wear impact-resistant gloves and safety goggles to prevent flying debris or ricochets from the impact components from causing injury;

Unauthorized personnel are prohibited from entering the test area to prevent accidental injury.

2.2 Environmental Safety

The ambient temperature should be maintained between 5°C and 35°C, with relative humidity ≤ 75%;

Leave at least 2000 mm of space in front of the equipment to ensure that the swing arm can lift properly and that there are no obstructions in the surrounding area;

When mounting the equipment on a wall, ensure it is vertical, with a tolerance of ≤0.5°.

3. Equipment Overview

3.1 Introduction to the Host

Introduction to the Host System Components (Figure 3-1).

Number Note
Mounting holes, 12 in total
Sample holder, with adjustable angle and position
Impact element, replaceable
Swing Arm
Adjustable arm mounting bracket
Fixed Device for the Impact Height of a Swing Arm
Display showing the current swing arm angle and impact height
Sensor that monitors the angle of the swing arm and automatically calculates the impact height

Pendulum Impact Test Device-Figure3

Figure 3-1

3.2 Replacing The Impact Element

Steps for Replacing the Impact Element (Fig. 3-2).

Number Note
Remove the two small screws at the top and bottom
Unscrew the nut
Unscrew the counterweight, then tap the connecting rod out.

Pendulum Impact Test Device-Figure4

Figure 3-2

The disassembled impact components are shown in the figure below (Figure 3-3):

Pendulum Impact Test Device-Figure5

Figure 3-3

Next, assemble the impact elements of different weights and install them onto the swing arm, paying attention to the orientation of the hammer head during installation.

3.3 Introduction to the Display

Display Introduction Features (Fig. 3-4).

Number Note
USB Charging Port
Display power switch: Press and hold for 1 second to turn on; press briefly to turn off.
Confirm Button/Reset Button
The adjustment button allows you to set the length of the swing arm; the factory default setting is 1 m.
Number Up/Down keys: Increase or decrease the number when setting parameters
Battery level indicator—please recharge promptly when the battery is low.
Current Impact Height Display
Current Swing Arm Angle Display

Pendulum Impact Test Device-Figure6

Figure 3-4

4. Impact Test

4.1 Sample Mounting and Installation

4.1.1 Sample Fixation

Secure the sample in the center of the plywood.

4.1.2 Position Adjustment

The position and angle of the sample holder are adjustable, and the position of the swing arm mounting device is adjustable. Adjust these so that the hammer head and the sample impact point are at the same height, and the direction of impact is perpendicular to the cross-section of the impact point.

4.2 Impact Parameter Settings

4.2.1 Set to 0 at Startup

When you turn on the display, it will show random angle and height values. Press the reset button to set the current swing arm angle and impact height to 0.

4.2.2 Adjusting the Impact Height

Raise the pendulum arm; the display will show its real-time angle and impact height. Refer to the table below to raise the pendulum arm to the desired impact height, and secure it using the pin on the height locking device. See Figure 4-1.

IK Level IK01 IK02 IK03 IK04 IK05 IK06
Energy J 0.14 0.2 (0.3) 0.35 (0.4) 0.5 0.7 1
Equivalent mass (kg) 0.25 (0.2) 0.25 (0.2) 0.25 (0.2) 0.25
Drop Height (mm) 56 (100) 80 (150) 140 (200) (250) 200 280 400

Note: The values in parentheses correspond to the requirements of the previous version of the standard. If testing is to be conducted in accordance with the previous version of the standard, a 150-gram impact element must be installed.

Pendulum Impact Test Device-Figure7

Figure 4-1

4.2.3 IK Impact Test

Slowly pull out the pin securing the swing arm height (Fig. 4-1) to perform the impact test.

To prevent the impact element from striking repeatedly due to inertia, after the impact element has completed an impact and bounced back, you must manually hold the impact element and slowly return it to its original position. Do not use your hand to directly grip the swing arm to stop it (as this can easily cause the swing arm to deform).

Adjust the position of the sample and continue conducting impact tests on the other test points until the entire test is complete.

4.3 Post-Test Procedures

4.3.1 Equipment Shutdown and Reset

Turn off the display and allow the impact element to hang freely.

4.3.2 Sample Preparation and Observation

Visual inspection: Check the sample housing for cracks, dents, or peeling, and check whether any fasteners are loose;

Functional Testing: When the sample is powered on, test whether all functions are operating normally and ensure there are no functional failures;

Determine whether the product meets the IK rating requirements based on the product standards, and record the testing process (test points, number of impacts, etc.) and the test results (pass/fail).

5. Daily Maintenance and Care of Equipment

5.1 Preparation of the Test Environment

Clean the equipment surfaces: Wipe down the test fixture, stage, and impact elements with a dry cloth to remove dust and grease. Do not use water or corrosive cleaning agents;

Tidy up the test area: Remove any sample debris (if present), organize tools and accessories, and keep the area clean and tidy.

5.2 Long-Term Storage and Maintenance (When Not in Use for More Than 1 Month)

Component Storage: After cleaning the impact components and swing arms, apply rust-preventive oil and place them separately in packaging boxes to prevent collisions;

Maintenance of Lubricated Parts: The pin on the swing arm height lock and the pivot on the swing arm lock must both be lubricated to ensure proper lubrication;

Equipment Protection: Cover with a dust cover; store in a dry, well-ventilated environment at a temperature of 5–35°C and humidity of ≤75%;

Charge regularly: Store the device after fully charging it, and charge it once a month to run a self-test.

5.3 Calibration Interval Requirements

To ensure the measurement accuracy and operational reliability of the equipment, it is recommended that it be sent periodically to a qualified third-party metrology laboratory for calibration. The recommended calibration interval is 12 months. Users may determine the specific calibration interval based on frequency of use, environmental conditions, and quality system requirements.

6. Troubleshooting and Resolving Common Problems

Fault Symptoms Possible Causes Procedure
The display won’t turn on Low battery on the display Try charging it again using the standard USB cable.
Abnormal altitude and angle readings Not reset to zero Every time the display is turned on, it shows random height and angle readings, and you must press the reset button to reset them.

7. Packing List

No. Name Quantity
IK01-06 Impact Element Test System Main Unit (includes a 150g impact element) 1 unit
Expansion Screw 12
USB Charging Cable 1 piece
200g shock element 1