Product No: LS-B65
The LS-B65 Frost Accumulation Water Evaporation Tester is designed to precisely simulate and measure frost accumulation in refrigerating appliances, fully compliant with IEC 60335-2-24 Clause 11.102 and Annex BB. It utilizes a controlled water evaporation method to generate frost under standardized conditions, ensuring accurate evaluation of defrosting system performance in refrigerators, freezers, and similar appliances.
Specifications:
1. Frost Generation System
• Heat Source:Built-in controllable heat source (precise power adjustment) to evaporate measured water, minimizing extraneous heat loss to the appliance cabinet.
• Structure:Integrates a thermal-insulating block enclosure (as per IEC 60335-2-24 Figure BB.1/BB.2 design) with a vertical central hole. Equipped with a high-thermal-conductivity evaporating dish base + low-thermal-conductivity walls, ensuring stable evaporation conditions.
2. Installation & Operation
• Positioning:Designed for mounting at the geometric center of the refrigerating appliance cabinet. Supports external electrical connection for voltage adjustment and power measurement with the cabinet door closed.
• Water Supply:Configured with a small-bore tube for interval-based water injection (no continuous flow required). Automatically calculates water volume based on cabinet gross volume (follows “2 g water per liter per week” standard rate).
3. Performance Parameters
• Evaporation Rate:Max ~2 g/h (at 4 W input, water at cabinet temperature), meeting standard test efficiency.
• Adjustability:Adapts to different defrost intervals (e.g., for a 140 L cabinet with twice-weekly defrost, supplies 140 g water as per IEC 60335-2-24 Annex BB NOTE).
4. Safety & Precision
• Energy Control:Ensures complete water evaporation without excessive energy input, avoiding cabinet overheating.
• Data Traceability:Records power, water volume, and evaporation time for compliance documentation.
IEC 60335-2-24 Figure BB.1
IEC 60335-2-24 Figure BB.2