Product No: LS-B60
The LS-B60 Circuit Breaker Tripping Characteristic Testing System is designed and manufactured in accordance with standards such as IEC 60947-1 Clause 7.2.1 and 7.2.2 (GB 14048.1-2012 Clause 7.2.1), IEC 60947-2 Clause 4.7 (GB 14048.2-2008 Clause 4.7), IEC 60898-1 Clause 9.10 (GB 10963.1-2005 Clause 9.10), and IEC 60009-1 Clause 3.7 (GB 16917.1-2014 Clause 3.7).
Specification:
Electrical Control Method | PLC intelligent integrated control, human-machine touch interface operation |
Test Chamber | The test chamber is combined with the test machine, with a built-in delay position and an instantaneous position, and is arranged on the back panel in a terminal manner to meet the 1-4P circuit breaker test. When the temperature inside the chamber is too low, it can be heated, ranging from room temperature to 50 degrees, with a temperature resolution of 0.1 degrees and a temperature stability of ±1 degree, providing a reliable and safe environment for circuit breaker testing. |
Test Items | • Small/molded case time-delay tripping characteristics: 1.13 times, 1.45 times, 2.55 times the current • Instantaneous tripping characteristics: 3~10 times the current • Temperature rise test |
Time-delay Tripping Test | |
Output Current | AC10~350A |
Output Voltage | 0~5V |
Output Current Path | 1-way output, multi-P circuit breaker contacts are tested by connecting several switches in series |
Test Sample Range | 3~125A |
Current Output Waveform | Sine wave (correction equipment current waveform distortion is no more than 5%, system error is no more than 3%) |
Current Control Method | Automatic adjustment, current setting on touch screen |
Current Accuracy | Class 0.5 |
Voltage Accuracy | Class 0.5 |
Transformer Ratio | 400/5A |
Testing Time | 1~9999H 59M 59s |
Testing Process | After applying 1.13 times the current, it will automatically rise to 1.45 times the current in 5 seconds |
Fixture | Adopting terminal mode, one delay position, one instantaneous position; meeting 1-4P circuit breaker |
Instantaneous Tripping Test | |
Output Current | AC100~1500A |
Output Voltage Number | 1 |
Current Accuracy | 0.5 Class |
Resolution | 1A |
Instantaneous Test Time | ≤0.2s |
Action Time Upper Limit Setting | 1~200ms |
Action Time Lower Limit Setting | 1~200ms |
Test Station | 1 |
Timekeeping | 1~9999ms |
Protection Function | Short circuit protection, overload protection, leakage protection |
Temperature Rise Test | |
Purpose of the Test | When the circuit breaker is tested under normal working conditions and specified conditions, the temperature rise of its components shall not exceed the limit values specified in Table 6, and the circuit breaker shall not be damaged to affect its function or safety of use. |
Test Current | AC10-350A |
Temperature Measurement Channel | 12 channels, temperature resolution 0.1 degrees |
Accuracy | 0.2 Class |
Measurement and Control Speed | 0.1 seconds per channel |
Temperature Measurement Range | 0~200℃ |
Thermocouple Model | T type |
IEC 60947-1 Clause 7.2.1 and 7.2.2
IEC 60947-2 Clause 4.7
IEC 60898-1 Clause 9.10