Product No: SMT-S15
The IEC 60065/IEC 60601 Standard Test Hook Probe is a critical pre-test tool for verifying enclosure integrity before accessibility testing. Crafted entirely from stainless steel for durability, it features a precision 90° hook design to secure into enclosure vents and seams. When pulled with a standard 20N force, it checks if enclosures can resist deformation—preventing unintended access to internal hazardous parts. Compliant with global safety standards, it’s widely used for audio/video devices and medical equipment, ensuring consistent and reliable pre-testing results.
Standards:
| Standards No. | Clause/Figure | Standards Name | 
| IEC 60065:2014 | Clause 9.1.7, Figure 4 | Audio, video and similar electronic apparatus – Safety requirements | 
| IEC 60601-1:2012/AMD1:2020 | Clause 8.1 | Medical electrical equipment – Part 1: General requirements for basic safety and essential performance | 
| UL 60065:2024 | Clause 9.1.7 | Standard for Safety of Audio/Video and Similar Electronic Apparatus | 
| CSA C22.2 No. 60065:2024 | Clause 9.1.7 | Audio, Video and Similar Electronic Apparatus – Safety Requirements | 
Specifications:
 Physical Dimensions:  Length: 180mm; Width: 5mm; Thickness: 1mm (all with ±0.1mm tolerance)
 Material:  304 stainless steel (corrosion-resistant, wear-proof for repeated use)
 Hook Design:  90° precision-bent hook (ensures secure attachment to enclosure seams/vents)
 Applied Force:  Designed for 20N pull force (standard for enclosure integrity pre-testing)
 Structural Feature:  Smooth, burr-free edges (prevents damage to enclosure surfaces during testing)
 Test Stage:  Pre-accessibility testing (prior to using test probes for hazardous part checks)
Test Procedures:
 1. Inspect the probe for bending, rust, or burrs; confirm dimensions and hook shape meet specs.
 2. Place the device (e.g., TV, medical monitor) on a stable, flat surface and power it off.
 3. Insert the probe’s hook into target enclosure areas: vents, seams, or gaps between panels.
 4. Apply a steady 20N pull force to the probe, holding for 5 seconds to simulate real-world stress.
 5. Check the enclosure for deformation, cracks, or loosened panels that could expose internal parts.
 6. Record results; if no damage occurs, proceed to formal accessibility testing.
Applications:
 1. Pre-accessibility testing of audio/video devices (e.g., TVs, speakers, DVD players) per IEC 60065:2014 Clause 9.1.7.
 2. Enclosure integrity checks for medical electrical equipment (e.g., patient monitors, ultrasound machines) following IEC 60601-1:2012/AMD1:2020.
 3. Quality control for consumer electronics (e.g., soundbars, projectors) during production to prevent weak enclosures.
 4. Compliance pre-testing for devices seeking UL 60065:2024 or CSA C22.2 No. 60065:2024 certification.
 5.Post-repair verification of medical device enclosures (e.g., replacement panels) to ensure safety compliance.

IEC 60065 / IEC 60601 Standard Test Hook Probe