Electrodynamic Vibration Generator System

User's Manual

Applicable LISUN models: LVD-100KG、LVD-100KG-6D

1. Basic Understanding of Equipment

1.1 Key Parameters

Specifications/Model Numbers LVD-100KG LVD-100KG-6D
Direction of Vibration Vertical and Horizontal Directions Vertical, horizontal, and longitudinal
Vibration Table Dimensions 500 mm × 500 mm (other sizes available upon request)
Vibration Waveform Sine Wave (Half-Wave, Full-Wave)
Vibration Control Fixed Frequency, Frequency Sweep, Programming

1.2 Equipment Configuration Verification

1.2.1 Unpacking

When unpacking the device, handle it gently to avoid scratching the casing with sharp objects. After unpacking, first inspect the exterior of the device to ensure there is no visible deformation or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.

1.2.2 Main Unit and Accessories

Host (Figure 1-1):

Number Note
Console
Vibration Table

Electrodynamic Vibration Generator System-Figure1

Figure 1-1

Main Components (Figs. 1–2):

Number Note
Handheld Vibration Meter and Accessories
Special mounting bracket for vibration tables, designed to prevent the vibration table from moving during operation
Strap For securing samples
Stud Clamping Device
Power Cord

Electrodynamic Vibration Generator System-Figure2

Figures 1–2

1.2.3 Downloading Electronic Documents

We have already sent the download links for the electronic user manual, warranty card, and other documents via email at the time of shipment. Please download them as soon as possible. If you missed the email, you can contact us to request the download links again.

2. Safety Precautions

2.1 Electrical Safety

Before connecting the power cord, verify that the supply voltage is AC 220 V ±10% and the frequency is 50 Hz (60 Hz is available upon request; please refer to your device’s nameplate); You must use a three-prong outlet with a protective ground, with a protective ground resistance of ≤4 Ω. Under no circumstances should you cut or disconnect the protective ground wire, as doing so poses a risk of electric shock.

2.2 Environmental Safety

The equipment must be installed in an air-conditioned room, with the ambient temperature maintained between 10°C and 35°C and relative humidity ≤ 80%;

The floor must be level, sturdy, and capable of supporting the required load; it must be free of flammable or explosive materials, heavy dust, and strong electromagnetic interference.

2.3 Operational Safety

The vibration test stand generates strong vibrations during operation; therefore, it is recommended to use it on a solid floor and secure it with a dedicated mounting bracket;

Do not pull, squeeze, or bend the cables; under no circumstances should power or signal cables be plugged in or unplugged while the device is running;

The sample must be securely clamped, with no looseness or displacement, to prevent splashes from injuring people or damaging the equipment;

Do not wear loose-fitting clothing or jewelry; tie back long hair; and wear safety goggles;

To ensure safety, it is recommended to install the equipment in a well-ventilated, separate testing area and to install protective barriers to prevent sample splatter;

When the equipment is in operation, personnel must remain at least 1 meter away from the work surface; reaching out or leaning toward the vibration zone is strictly prohibited;

If you notice unusual noises, odors, smoke, loose samples, or an overload, immediately shut down the control console, troubleshoot the issue, and then try again.

2.4 Other

For matters not covered in this manual, please proceed with caution or contact us.

3. Equipment Overview and Installation

3.1 Introduction to the Console

See Figure 3-1.

Number Note
Power Button
Vertical Vibration Indicator Light
Lateral Vibration Indicator Light
Vertical Vibration Indicator Light
Touchscreen

Electrodynamic Vibration Generator System-Figure3

Figure 3-1

3.2 Introduction to and Installation of the Vibration Test Stand

3.2.1 Introduction to the Vibration Test Stand

See Figure 3-2.

Number Note
M8 screw holes, for securing samples with screws
Stud holes, which can be used to secure samples with a stud clamping device

Electrodynamic Vibration Generator System-Figure4

Figure 3-2

3.2.2 Vibration Table Installation

Secure the vibration table’s mounting bracket to a level surface using expansion screws, then place the four legs of the vibration table securely into the mounting bracket.

3.3 Wiring Between the Control Console and the Vibration Table

See Figure 3-3.

Number Note
The vibration test stand includes two cables.
Connect the two cables from the vibration table to the control console, pairing them one-to-one.
Power cord jack, for connecting the device to a suitable power source

Electrodynamic Vibration Generator System-Figure5

Figure 3-3

4. Preparations

4.1 Sample Fixation

Secure the specimen as described in Section 3.2.1 and mount it as closely as possible to the vibration table to ensure that no relative displacement occurs between the specimen and the vibration table during the test.

4.2 Introduction to Handheld Vibration Meters

4.2.1 The Function of a Vibration Meter

Since different samples have different weights and centers of gravity, the amplitude and acceleration will vary even when the same vibration intensity is used. Therefore, the test program cannot directly set the amplitude and acceleration; instead, the required vibration intensity must be adjusted using a vibration meter based on the specific sample and known amplitude or acceleration values.

4.2.2 Assembly of the Vibration Meter

See Figure 4-1.

Number Note
Vibration Meter
Connect the vibration meter and the magnet using a signal cable.
Magnet
Remove the broken magnetic strip from the magnet

Electrodynamic Vibration Generator System-Figure6

Figure 4-1

5. Vibration Testing

5.1 Introduction to the Touchscreen Program

After turning on the device, tap the touchscreen to access the initial screen, then tap “Mode Selection,” as shown in Figure 5-1.

Electrodynamic Vibration Generator System-Figure7

Figure 5-1

Access the test mode selection screen (see Figure 5-2). Click the desired test mode to enter the parameter settings and test screen.

Number Note
Vertical Fixed Frequency
Vertical Sweep
Horizontal Fixed Frequency
Horizontal Sweep
Vertical Fixed Frequency
Vertical Sweep
Programmable Fixed Frequency
Programmable Sweep

Electrodynamic Vibration Generator System-Figure8

Figure 5-2

Note: “Vertical” refers to vibration in the vertical direction; “horizontal” refers to vibration in the horizontal direction; and “longitudinal” refers to vibration in the longitudinal direction.

5.2 Fixed-Frequency Vibration Testing

5.2.1 Introduction to the User Interface

The interfaces and operations for vertical fixed-frequency, horizontal fixed-frequency, and longitudinal fixed-frequency are identical. Taking vertical fixed-frequency as an example, click to enter the vertical fixed-frequency interface, as shown in Figure 5-3.

Number Note
Displays the duration of the test so far, in seconds
Displays the remaining test time in seconds
Displays the current vibration frequency in Hz
Displays the current vibration intensity, in percent
Current: Displays the current operating status of the device, in amperes (A)
Set the test duration, in seconds
Set the vibration frequency in Hz
Set the vibration intensity, in %
Acceleration slope, in percent. For fixed-value testing, the default acceleration slope is 2%; no modification is necessary.
Click to start the test
Click to stop the test
Click to toggle between full-wave and half-wave
Click to clear test data
Click to view the test curve

Electrodynamic Vibration Generator System-Figure9

Figure 5-3

5.2.2 Vibration Intensity Calibration

Attach the vibration meter’s magnet to a surface parallel to the direction of vibration. For vertical vibration, attach it directly above the vibration source. For horizontal vibration, attach it to a side parallel to the direction of vibration. Turn on the vibration meter and set it to the displacement mode (in mm) to measure the amplitude of the vibration table; alternatively, set it to the acceleration mode (in m/s²) to measure the vibration acceleration of the vibration table.

Note: Vibration test standards require either amplitude or vibration acceleration, but not both.

On the console’s touchscreen interface, you can first enter a longer test duration, then enter the desired vibration frequency, and start the test after entering a lower vibration intensity—for example, 10%.

Once the current vibration intensity reaches the set value, check the amplitude and vibration acceleration displayed on the vibration meter, as shown in Figure 5-4.

Electrodynamic Vibration Generator System-Figure10

Figure 5-4

Then, based on the difference between the measured values and the standard requirements, adjust the vibration intensity until the standard requirements are met.

Note: For the same frequency and vibration intensity, the amplitude and acceleration produced in the three vibration directions are not the same. To facilitate testing, the orientation of the sample may be adjusted as needed.

5.2.3 Startup Test

Once you have determined the required vibration intensity, re-enter the desired test duration and vibration intensity, and then click “Start” to begin the actual test.

5.3 Sweep Vibration Testing

5.3.1 Introduction to the User Interface

The interface and operation for vertical sweep, horizontal sweep, and longitudinal sweep are the same. Taking the vertical sweep as an example, click to enter the vertical sweep interface, as shown in Figure 5-5.

Number Note
Set the total test duration
Set the frequency slope (e.g., if the high-frequency value is 20 Hz and the low-frequency value is 10 Hz, and the sweep must be completed within 5 seconds, then the frequency slope = (20 Hz – 10 Hz) / 5 s = 2 Hz/s)
Set the acceleration slope (e.g., if the high-frequency vibration intensity is 20% and the low-frequency vibration intensity is 10%, and the sweep must be completed within 5 seconds, then the acceleration slope = (20% – 10%) / 5 seconds = 2%/s)
Set the low-frequency test frequency
Set the low-frequency test vibration intensity
Set the low-frequency test duration
Set the high-frequency test frequency
Set the high-frequency test vibration intensity
Set the duration of the high-frequency test

Electrodynamic Vibration Generator System-Figure11

Figure 5-5

Note: Changes in vibration intensity may not be linear. For example, when conducting a sweep test from 10 Hz to 55 Hz at an amplitude of 0.35 mm, to avoid anomalies during the test, the 10 Hz to 55 Hz range can be divided into multiple segments.using the method described in Section 5.2.2 to record the vibration intensity required for each frequency point to reach an amplitude of 0.35 mm. The frequency range should then be divided into multiple segments, and a programmable sweep test should be conducted in accordance with Section 5.5.

5.3.2 Startup Test

Once all parameters are set, simply click “Start” to begin the official test.

The vibration test stand first enters the low-frequency test. Once the low-frequency test duration has elapsed, it performs a frequency sweep to the high-frequency test according to the set frequency ramp and vibration intensity ramp; Once the high-frequency test duration has elapsed, the system performs a frequency sweep back to the low-frequency test range according to the preset frequency and vibration intensity ramp rates; this process continues until the total test duration reaches the preset value.

5.4 Programmable Fixed-Frequency Vibration Testing

5.4.1 Introduction to the User Interface

To access the programmable fixed-frequency settings interface, refer to Figure 5-5 for the parameters to be configured.

Number Note
Vertical orientation only is supported. Please mount the sample in the vertical direction on the vibration test stand. Switch the corresponding segment from “OFF” to “ON.”
Paragraph Number
Set a different vibration frequency for each section
Set a different vibration intensity for each segment
Set a different test duration for each section
Acceleration slope: Use the default value of 2%; no modification is necessary.
Time interval between two consecutive test segments
Number of cycles
Program Number
Save the program, or call the program
Click the right arrow to edit paragraphs 4, 5, and 5.
Click the left arrow to enter the test interface.

Electrodynamic Vibration Generator System-Figure12

Figures 5–6

5.4.2 Startup Test

Once all parameters are set, simply click “Start” to begin the official test.

The vibration test stand first enters Test Segment 1. Once the test duration reaches the set interval, it proceeds to Test Segment 2. After completing all test segments, one cycle is completed; this process continues until all cycle tests are completed according to the set number of cycles.

5.5 Programmable Sweep Vibration Testing

5.5.1 Introduction to the User Interface

To access the programmable sweep settings interface, refer to Figure 5-7 for the parameters to be configured.

Number Note
Vertical orientation only is supported. Please mount the sample in the vertical direction relative to the vibration table for testing. Switch the corresponding segment from OFF to ON.
Segment number. The following parameters must be set in order for this segment: low-frequency vibration frequency, low-frequency vibration intensity, high-frequency vibration frequency, high-frequency vibration intensity, sweep test duration, frequency ramp rate, acceleration ramp rate, low-frequency test duration, and high-frequency test duration.
Time interval between two consecutive test segments
Number of repetitions
Program Number
Saving and Loading Programs

Electrodynamic Vibration Generator System-Figure13

Figure 5-7

5.5.2 Startup Test

Once all parameters are set, simply click “Start” to begin the official test.

The vibration table first enters the first segment of the low-frequency test. Once the low-frequency test duration ends, it performs a frequency sweep up to the high-frequency test according to the set frequency slope and vibration intensity slope. Once the high-frequency test period ends, the system proceeds to the second low-frequency test segment, continuing in this manner until all segments of the sweep test are completed, thus finishing one cycle. All cycle tests are completed according to the set number of cycles.

5.6 End of Testing

5.6.1 Stopping the Test

If an anomaly occurs during the test, please stop the test immediately, troubleshoot the issue, and then restart the test;

If everything goes smoothly during the test, wait for the test to complete normally.

5.6.2 Inspecting Samples

After the test is complete, inspect the sample’s mechanical and electrical performance to determine whether it has passed the test.

6. Daily Maintenance and Care of Equipment

6.1 Daily Maintenance

Check that the work surface, control panel, and cables are free of damage, oil stains, and loose connections;

Wipe down the countertop and touchscreen, and remove any debris;

Check the screws securing the vibration table and make sure none are loose;

The tabletop operates smoothly, with no unusual noises or jamming.

6.2 Storage Specifications

6.2.1 Storage Environment Requirements

Temperature: 10°C to 35°C; relative humidity: ≤80% RH; dry, dust-free environment free of corrosive gases.

6.2.2 Long-Term Storage Operations

Turn off the power and unplug all cables; clean the equipment and wrap it in dust-proof film; power on the vibration test stand for 10 minutes each month to prevent moisture damage to the components.

6.3 Calibration Interval Requirements

To ensure the measurement accuracy and operational reliability of the equipment, it is recommended that it be sent periodically to a qualified third-party metrology laboratory for calibration. The recommended calibration interval is 12 months. Users may determine the specific calibration interval based on frequency of use, environmental conditions, and quality system requirements.

7. Troubleshooting and Resolving Common Problems

Fault Symptoms Possible Causes Procedure
The sample amplitude is significantly too large. The sample is not secured, and its center of gravity is too far from the vibration table Refer to Sections 4.1 and 3.2.1. Secure the specimen to the vibration table and ensure that no relative displacement occurs between the specimen and the vibration table during the test. Secure the specimen as closely as possible to the vibration table so that its center of gravity is near the table surface.
Touchscreen Overload Protection Alert Sample overweight; vibration table operating beyond capacity Please stop the test, replace the sample with a lighter one, and try again.

Appendix & Downloads