RF Conducted Immunity Test System
User's Manual
Applicable LISUN models: AB-EM
The AB-EM electromagnetic clamp complies with the national standard IEC 61000-4-6 and is a specialized product for testing the immunity of equipment to radio-frequency field-induced conducted disturbances.injecting disturbance signals into the cable assemblies in the slots so that the Equipment Under Test (EUT) is subjected to the levels specified in the national standard.
For precise measurements, the electromagnetic clamp is also equipped with an auxiliary absorption clamp and calibration pieces.
Features
See Tables 1 and 2 for the coupling coefficient and attenuation characteristic curves; the difference is 9.5 dB ± 1.5 dB.
Maximum input power: 100 W (5 minutes);
Injection frequency: 150 kHz to 230 MHz;
At frequencies greater than 10 MHz, the coupling/decoupling directivity is greater than 10 dB
Measurable cable diameter: <Φ18 mm
Attachments
Instruction Manual 1 copy
Test Cable One
100W 6dB Attenuator One
Calibration Fixture (Optional) One set
Using Connection Diagrams
AB-EM

AB-EM Data Sheet
Serial No.: 21021 Differential Error: 9.5 dB ± 1.5 dB Date: April 29, 2021
| Frequency (MHz) | Coupling Coefficient (dB) | Attenuation Coefficient (dB) | Difference (dB) |
| 0.15 | -2.7 | -12.3 | 9.6 |
| 0.2 | -2.4 | -12.0 | 9.6 |
| 0.3 | -2.2 | -11.8 | 9.6 |
| 0.32 | -2.2 | -11.8 | 9.6 |
| 0.35 | -2.2 | -11.8 | 9.6 |
| 0.4 | -2.1 | -11.8 | 9.7 |
| 0.45 | -2.1 | -11.8 | 9.7 |
| 0.5 | -2.1 | -11.8 | 9.7 |
| 0.55 | -2.1 | -11.8 | 9.7 |
| 0.6 | -2.1 | -11.8 | 9.7 |
| 0.65 | -2.1 | -11.8 | 9.7 |
| 0.74 | -2.1 | -11.8 | 9.7 |
| 0.75 | -2.1 | -11.8 | 9.7 |
| 0.8 | -2.2 | -11.9 | 9.7 |
| 0.85 | -2.3 | -11.9 | 9.6 |
| 0.9 | -2.4 | -12.1 | 9.7 |
| 0.95 | -2.5 | -12.1 | 9.6 |
| 1 | -2.5 | -12.1 | 9.6 |
| 1.5 | -2.1 | -11.7 | 9.6 |
| 2 | -2.1 | -11.7 | 9.6 |
| 2.5 | -2.1 | -11.7 | 9.6 |
| 3 | -2.1 | -11.7 | 9.6 |
| 4 | -2.2 | -11.8 | 9.6 |
| 5 | -2.2 | -11.8 | 9.6 |
| 6 | -2.3 | -11.8 | 9.5 |
| 7 | -2.4 | -11.8 | 9.4 |
| 8 | -2.4 | -11.8 | 9.4 |
| 9 | -2.5 | -11.9 | 9.4 |
| 10 | -2.6 | -11.9 | 9.3 |
| 12 | -2.8 | -12.1 | 9.3 |
| 15 | -3.2 | -12.4 | 9.2 |
| 18 | -3.5 | -12.6 | 9.1 |
| 20 | -3.6 | -12.8 | 9.2 |
| 22 | -3.6 | -12.8 | 9.2 |
| 25 | -3.5 | -12.8 | 9.3 |
| 28 | -3.1 | -12.4 | 9.3 |
| 30 | -2.6 | -11.9 | 9.3 |
| 32 | -2.2 | -11.5 | 9.3 |
| 35 | -1.3 | -10.6 | 9.3 |
| 38 | -0.5 | -9.8 | 9.3 |
| 40 | 0.0 | -9.3 | 9.3 |
| 45 | 1.1 | -8.4 | 9.5 |
| 50 | 1.8 | -7.8 | 9.6 |
| 60 | 2.5 | -7.4 | 9.9 |
| 70 | 2.6 | -7.1 | 9.7 |
| 80 | 2.7 | -6.9 | 9.6 |
| 85 | 2.9 | -6.9 | 9.8 |
| 90 | 3.0 | -6.9 | 9.9 |
| 95 | 3.2 | -6.9 | 10.1 |
| 100 | 3.3 | -6.8 | 10.1 |
| 120 | 3.5 | -6.8 | 10.3 |
| 140 | 3.4 | -7.0 | 10.4 |
| 160 | 2.6 | -8.0 | 10.6 |
| 180 | 1.9 | -8.2 | 10.1 |
| 200 | 0.9 | -9.3 | 10.2 |
| 210 | 0.6 | -9.6 | 10.2 |
| 220 | 0.1 | -10.1 | 10.2 |
| 230 | -0.1 | -10.3 | 10.2 |


List of AB-EM Calibration Fixtures
| No. | Model | Name | Quantity/piece |
| 1 | M-004 | 150Ω/50Ω Adapter | 2 |
| 2 | M-006 | 5 W / 50 Ω load | 1 |
| 3 | E-001 | L-Shaped Calibration Bracket | 2 |
| 4 | E-003 | 4mm Test Rod (725mm) | 1 |
*The 4mm test probe is inside the electromagnetic clamp packaging box;
LED Luminaires and LED Power Driver Test Solution
Goniophotometer System: LSG-6000, LSG-1890B, or LSG-1700
Spectroradiometer & Integrating Sphere Test System: LPCE-2 or LPCE-3
LED Life Maintains a Test System in Accordance with LM-80: LEDLM-80PL
Waterproof Test for IPX5 and IPX6 Levels: JL-56
Dustproof Testing Machine for IP5X and IP6X Testing: SC-015
LED Power Driver Testers: LS2090 (for the lab) and ATE-2 (for the production line)
Electrical Safety Tester: LS9955
CFL and Electronic Ballast Test Solution
Goniophotometer System: LSG-1890B or LSG-1700
Integrating Sphere and Spectrophotometer Test System: LPCE-2 or LPCE-3
Adjustable Reference Ballast: DYJ-50HZ, HCS-109A, and DYJ-HID
Electric Ballast Tester: WT5000 (for the lab) and ATE-1 (for the production line)
Digital Torsion Meter and Multiway Life Tester: CH338 and CH316
Electrical Safety Testers: LS9923, ZRS-3H, ZY-3
EMC and EMI Test Solutions for CFL and LED Luminaires
EMI Test System: EMI-9KB or EMI-9KA
Electrostatic Discharge Simulator: ESD61000-2
EFT Immunity Measurement: EFT61000-4
Surge Generator: SG61000-5
Voltage Dips and Interruptions Generator: CSS61000-11
Ring Wave Generator: RWG61000-12

www. Lisungroup.com
Patents Filed
All rights reserved by Lisun Group
Subject to Change Without Notice

中文简体